Reloading CM200 or CM120 software
Courtesy of Bob Murphy, FEI
All CM200feg and CM120 Users:
From time to time the software on both the CM200feg and CM120 will probably
need reloading. With all the power glitches and problems this year the
problems will get slightly worse. You will know you will need to reload it
if you see the message "Waiting for PC to attach". These 2 instruments have
a feature called "Downloadable Lens Series" which allows you to edit and
change lens values. The problem is that because of this option the memory is
RAM which is more Volatile then Rom memory. Any power glitch can cause even
1 bit to change value which will make it necessary to reload all the memory.
The following procedure needs to be used.
- Turn the power off and on to verify that a memory block is failing
consistently.
- After the CM does some selftests you will get the message at the bottom
"Waiting for PC to Attach".
- On your laptops close windows and turn the laptop completely off (this is necessary to reset the COM port properly). Turn the laptop back on by
holding the on button for a few seconds. After windows loads verify the
cable to the Laptop is connected well and the switch box is set to PC.
- Open to downloadable lens group. Double click the download lens Icon. The program will start-up. Click "select file" and in the list select "122.exe"
for the CM200feg and "112.exe" for the CM120. These are the versions of the
software you should be running then Click OK.
- Click the download button and the Laptop should show download in progress and the bottom of the CM monitor should show Download in progress. On the CM
monitor a number will count up slowly to about 80. NOTE: On occasion there
is a timing issue so if the procedure doesn't work try getting the Laptop
ready and hitting download even before the CM says "Waiting for PC to
Attach". You can do this as the CM is testing blocks 5 or 6.
- After the Loading is done the CM will restart and should pass all the
tests. After the tests you can click on the laptop "Start CM". The
microscope will start-up normally and the vacuum system will need to warm
up.
- To speed things up especially if wet film was just introduced before the
glitch you can remove the film. You will also have to remove the Holder
since after the CM starts it must reset it and do a home. If the
anticontaminator is cold the system will also restart slower because it will
automatically vent and could introduce a little moisture when air hits the
cold blades.
NOTE: If you go to the configuration page on the CM and find there are no
Aperture sizes listed then you will also need to run CMonitor from the
laptop and restore an alignment to the microscope. You will also need to put
the film information in like Emulsion, Data Intensity, Stock and exposure
numbers and also redue the apertures sizes again. On the CM12O you will need
to do a Gun Alignment Procedure and filament limit. Unfortunately these must
be resored manually. You can find what data and numbers to put in by looking
for PM sheets on the shelves. DO NOT do anything if the aperture sizes are
listed which means the alignment memory is good.